<aside> ⚠️

Important reminders

Maintenance schedule

Action Frequency
Check water level in humidity pan, 1/2" from top Weekly
Change HEPA and gas filters 6 months
Inspect door latch hinges and door gasket seal Yearly
Verify and document CO₂, humidity, temp calibration Yearly
Replenish rust inhibitor in water jacket 2 years
Perform complete decontamination procedure As needed

Filling the humidity pan

  1. Use sterilized, distilled, demineralized or deionized water. Water must be safe with stainless steel — no high purity, ultra pure or milli-Q water. The pan is autoclavable.
  2. Fill the pan to within 1/2 inch of the top with sterile, distilled water. The pan requires approximately 3 L of water to fill to within 1/2 inch of the top.

Calibration

See "Section 2: Calibration" in the manual.

Cleaning and decontamination

No bleach or any disinfectant with high chlorine content. See "Section 5: Routine Maintenance" in the manual for thorough cleaning instructions.

Changing HEPA and gas filters

See "Sections 5–6" in the manual for instructions. An alarm can also be set to alert personnel when replacements need to occur (Section 5-6).

Rust inhibitor replenishment

  1. Mix 1 bag or bottle of rust inhibitor with a gallon of distilled, demineralized, or deionized water.
  2. Drain a gallon of water from the jacket and replace it with the rust inhibitor mixture.