Locator Jr. Plus specifications
- Liquid N₂ capacity: 71 L
- Liquid N₂ evaporation rate: 0.85 L/day
<aside>
🚨
Warnings
- The tank uses a vacuum to ensure temperature control, and care must be taken to prevent premature vacuum loss.
- Do not tamper with or remove the vacuum port (covered by a black plastic cap on the side of the vessel).
- When inserting or removing racks, be careful not to come into contact with the neck tube area of the vessel. Remove them slowly, in a vertical manner. Scratches on the neck tube area can cause premature vacuum failure.
- Do not use the tank for transport. Use an intermediate liquid N₂ vessel to move liquid N₂ from the dock to the tank. Cracks in the floor and bumping into objects can disrupt the vacuum.
</aside>
Adding liquid N₂ to a warm vessel
- Add only a small amount of liquid N₂ (5–10 liters) to the new or warm vessel. Liquid N₂ is available on the first floor on the dock.
- Allow this small amount to sit in the covered vessel for at least 2 hours, to let the vessel equilibrate to the temperature change.
- Add an additional 15 liters and allow the vessel to sit for 48 hours. Monitor liquid N₂ consumption with a wooden yardstick — never a hollow tube or plastic dipstick.
- Fill the vessel no higher than the top of the racks, approximately 12 inches with racks inserted.
- Insert and remove racks slowly. Allow the liquid N₂ to run out of boxes and off racks before completely removing them from the Locator.
Maintenance
- Monitor the level of liquid N₂ in the tank every 1–2 days to anticipate subtle changes in consumption and prevent problems before they arise. Keep a record of the levels using the tank level log below.
- No vacuum will last forever, so have the vessel tested by the manufacturer (Thermo Fisher) or a qualified cryo-vessel service technician every 7–10 years.
Liquid N₂ Tank Level Log
Liquid N₂ Tank Level Log