Retrieving CO₂ tanks
- Order "Industrial Grade Carbon Dioxide, 50 Pound Cylinder, CGA-320" from Airgas (Airgas Part # CD 50) via MDS.
- CO₂ cylinders are chained to the wall in the dock of Hanson. Use a gurney to safely transport the cylinder to the elevator, up to the second floor, to the incubators in Room 223.
- Chain the CO₂ cylinder to the wall next to the incubators.
<aside>
⛓️
Do not forget to chain the remaining cylinders back to the wall after removing yours.
</aside>
Connecting the tank to the pressure regulator
- Attach the regulator to the cylinder using the small white washer (Part # RAD64003940) to create a seal between the regulator and the cylinder. Tighten the inlet nut securely with pliers.
- Close the regulator by turning the adjusting knob (the one with the big arrow) to the full counter-clockwise position. The regulator must be closed before opening the cylinder valve.
- Open the cylinder valve slowly to allow the pressure to rise gradually in the regulator. When the high-pressure gauge stops rising, open the cylinder valve fully.
- Adjust the pressure by turning the adjusting knob clockwise until the low-pressure gauge shows the required output pressure. Input pressure to the incubators must be maintained at 15 psi.
- Ensure there are no leaks in the connection between the regulator and the cylinder by pipetting ~100 μL of soapy water around the connection. If bubbles start forming, there is a leak — close the cylinder valve and unscrew the inlet connection, then try forming a tighter seal with the washer and the regulator connection.
- The incubator has serrated fittings on the back of the cabinet to connect the gas supply. Ensure the regulator has tubing running to the fitting labeled "CO₂ inlet #1 tank".
Removing a tank from the regulator
- Close the cylinder valve to isolate the regulator from the pressure source.
- Vent the gases in the regulator by turning the adjusting knob clockwise and noting the pressure drop on the output gauge.
- After relieving the gas pressure, turn the adjusting knob counter-clockwise as far as possible.
- Disconnect and remove the cylinder from the regulator.